Cleaning methods for quartz test tubes
Leave a message
Conventional cleaning methods
Generally speaking, for every 2-4 samples analyzed, the quartz combustion tube (which can be likened to a quartz test tube) needs to be cleaned. If there is a lot of dust, the cleaning frequency needs to be increased. When the analysis time is too long and the accumulated splashes on the tube wall are difficult to remove by the wheel brush, it is necessary to remove the quartz test tube for cleaning. The specific steps are as follows:
Preparation: Press the up and down button on the analyzer panel to lower the P&T tray; Turn off the power gas source.
Dismantling components: Remove the O-ring seal at the bottom of the quartz test tube; Remove the gas path pipe from the furnace head cleaning component; Loosen the two compression nuts of the furnace head cleaning component; Remove the furnace head cleaning component; Remove the quartz test tube.
Soaking treatment: Soak the quartz test tube in a (1: 3) hydrochloric acid solution heated to 60 ℃ -80 ℃. After the splashes on the surface of the test tube are basically removed, take out the test tube and rinse it with clean water.
Remove residual dirt: Use a test tube brush to brush away residual dirt.
Drying treatment: Place the test tube in an oven to dry.
Installation and inspection: Install the test tube, furnace head cleaning component, and O-ring at the bottom of the test tube (apply a thin layer of vacuum silicone grease on the inner wall of the sealing ring); Perform a leak check.
Professional cleaning machine cleaning method
Quartz tube cleaning machine cleaning
When cleaning, pay attention to wearing protective goggles and acid resistant gloves. Place the quartz test tube to be cleaned into the groove inside the quartz tube cleaning machine, with the tube opening facing the velvet making room and the tube tail facing the diffusion furnace. Pull down the protective door of the machine, set the cleaning time to 30 minutes, and finally press the start button to start cleaning. After the cleaning time is up, put on goggles and acid resistant gloves, take out the quartz test tube from the acid tank, and gently place it in the sink. After the water in the sink is drained, rinse the inside and outside of the quartz test tube with a water gun and spray evenly for about 4 minutes. Do not press the start button when rinsing the inside of the test tube. If it is in the start state, immediately turn it off. When cleaning the test tube mouth with a water gun, rinse it from the inside to the tail direction, and operate carefully to ensure that it is clean and leaves no residue. Rotate the quartz test tube 180 degrees every two minutes, rinse it, and then rotate it again until it is completely cleaned. Add water to the sink until the water level reaches the middle of the test tube tail. The cleaning time for the sink is about 30 minutes. After the time is up, drain the water in the sink and repeat the flushing operation to ensure that there are no residual particles. Finally, wrap the mouth and tail of the pipe with foam paper and seal them with aluminum foil. Newer quartz test tubes only need to be cleaned for about 30 minutes, while used quartz test tubes need to completely remove the white pentoxide attached to the inner wall of the tube. Some may need to be cleaned for one or two days. When the acid tank machine is not started, rotate the quartz test tube 180 degrees every two hours to clean it more evenly and cleanly.